BodyBox New Calibration Method

Modified on Fri, 4 Apr at 8:39 AM

For 2024 upwards Bodystik devices, auto calibration can be enabled in the device information settings. 


the differences:


  • It's first creating a positive pressure (with the auto calibration pump) in the cabin to determine the positive time constant
  • Then it creates a negative pressure to determine the negative time constant
  • The final time constant is then the average of both time constants


and the advantages:


  • It is always calibrated with the same pressure and it does not depend on how hard the user closes the door to build up the pressure
  • It is less affected by thermal disequilibrium