Flow sensor Deviation Warning

Modified on Wed, 23 Oct at 12:00 PM

If you see a warning message like this pop up after a calibration:


In the newest update of Blue Cherry 1.3.4.1 it takes the most recent calibration and checks the average against past recent calibrations so if there is a slight change in the average of the calibrations then it will produce this warning message.


To help prevent this message from popping up you need to enable flow sensor selection here is a guide on how to do this: https://lovemedical.freshdesk.com/en/support/solutions/articles/48001245194-logging-flow-sensors


If this still pops up even with selection available make sure you have selected the correct flow sensor in the selection menu. If you have selected the correct flow sensor and the warning message is stating that there could be some changes to the properties of the flow sensor and therefore the calibration average would have changed. If the average stays the same after this warning message then the warning will eventually go away as the average is staying consistent. It is fine, as a warning is still a pass but it can mean the flow sensor is starting to age/deteriorate.


Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons

Feedback sent

We appreciate your effort and will try to fix the article